XMM-Newton Phase II RPS Users Manual
Up: A. Observation examples
Previous: A..3 Imaging observation of a point source
A..4 Timing observation of a variable source
Consider observing a bright point source, as above (like, e.g., an unresolved
AGN, binary or stellar object), but now with a special interest in high
time resolution measurements.
The input of standard information (target name, catalogued position etc.)
in XRPS is trivial. Other input parameters require some more thought:
- Choice of prime instrument
- The prime instrument is chosen according to the importance of data from
either type of XMM-Newton instrument: either EPIC, if imaging with moderate
resolution spectroscopy is crucial, or RGS, if the highest possible
spectral resolution must be achieved.
Let us, for the time being, assume that EPIC is prime.
- Science mode of the prime instrument
- For high time-resolution observations the EPIC pn TIMING mode is a
suitable choice.
- X-ray properties of the source
- Based on PIMMS observers can convert ROSAT, ASCA or other known
flux and band data (if known), and enter these, together with an
X-ray spectral model, e.g. power law, = 0.7, N(H) = 3e21
cm and the lower and upper limit of the energy band over which
the X-ray flux was observed.
- Duration of observation vs. visibility constraints
- Users must check that the requested observation fits into a continuous
visibility period of the XMM-Newton orbit, using the XMM-Newton Target Visibility Tool. In case that the
required total integration time is longer than the longest possible
visibility window, the observation must be split into an adequate
number of individual observations.
- Pointing coordinates
- No boresight coordinates need to be entered to centre the target on the
prime instrument, because the target coordinates will be propagated
automatically into the boresight fields, if no other values are
provided. The best data quality will be achieved in the aim point of
the prime instrument.
- Science modes of the other instruments
- Based on the nature of the example, one can assume that all instruments
will be operated in their fast modes. EPIC pn reaches a time resolution
of 0.03 ms in its TIMING mode, MOS reaches a resolution of 1.5 ms. Note
that the pn camera in its Small Window mode already reaches a time
resolution of 6 ms, which would at the same time render possible imaging
of the target. RGS would be operated in the SPECTROSCOPY mode and OM in
the Science User Defined mode with Image and Fast windows.
- Avoidance of nearby bright sources
- Optical and X-ray catalogues should be searched for nearby bright
sources which might lead to contamination of either the X-ray (e.g., RGS spectral overlaps) and/or optical/UV observations. In particular, when
the EPIC TIMING mode is used, one must ensure that nearby sources do not
contaminate the target data. There must be no nearby source in the same
column of the EPIC cameras as the science target. However, note that the
two MOS cameras are mounted orthogonally to each other. For the RGS there must be no source along the dispersion direction of the target
spectrum. Such sources must be avoided, which might lead to a position
angle constraint.
- EPIC filters
- Using the instructions provided in the XMM-Newton Users Handbook on EPIC filters
, the user
must decide which optical blocking filter suppresses optical loading in
the soft part of the X-ray passband sufficiently and at the same time
has minimal impact on the proposed science.
- RGS readout sequence
- The user must determine whether one or several CCDs shall be read
out. The whole spectral range (all 8 CCDs) can be read out in 5 s for
RGS1 and 10 s for RGS2.
- OM brightness limit
- Before planning details of OM observations, users should check for the
presence of bright optical/UV sources within the OM's FOV. There should
be no source in the FOV that violates the brightness constraints tabulated
in UHB Table 25
. If such source should exist,
no OM exposure should be included.
- OM filters and modes
- OM filters and modes have to be chosen according to the optical characteristics of the target
(see 5.2.4.5).
- Duration of exposures
- Normally, X-ray observations can be obtained in a single
exposure covering the entire duration of the observation.
The OM exposure times should be chosen according to the explanations
in § 5.2.4.5 and the OM chapter
of the XMM-Newton Users Handbook.
Up: A. Observation examples
Previous: A..3 Imaging observation of a point source
European Space Agency - XMM-Newton Science Operations Centre